Dwyer UFC User Manual Page 21

  • Download
  • Add to my manuals
  • Print
  • Page
    / 56
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 20
3: Operating Procedures
17
3.1 Setting-up the Instrument
3.1.1 Using the instrument for the first time
Initial user language selection
The first time you power-up the instrument you will be asked to select a user language, which will then be the
default language when the instrument is next used. If you want to change the language when the instrument is in
use, see below.
1. On initial power-up, the start-up screen will be
displayed for 5 seconds, showing the
instrument's serial number and software
revision.
2. After 5 seconds, the available language list will
be displayed.
3. Select the required language and press ENTER.
4. The instrument will display the MAIN MENU.
The MAIN MENU screen
The MAIN MENU screen is at the top of the menu
hierarchy and is the starting point for all the
operations described in this chapter. This screen is
normally accessed from the FLOW READING screen
by pressing the ENTER key.
Setting the date & time
5. Select Setup Instrument from the MAIN
MENU and press ENTER. The SETUP
INSTRUMENT screen should now be displayed.
6. Select Set Date & Time then press ENTER.
7. A flashing cursor should appear under the first
date number. Enter the date sequence in
dd-mm-yy order then press ENTER.
8. Repeat this action to set the time.
9. Select Exit then press ENTER to return to the MAIN MENU.
Note: If you make a mistake when entering the data press the Delete key to move the cursor back to the
number you wish to change, then continue. If you enter an invalid number an ‘ERR:Invalid Date or Time!’
error message is displayed on the second line of the screen. If this occurs repeat the set date/time procedure.
Key Point: When the instrument used for the first time the operator has free access to
all the set-up and operating menus until the instrument is put into FLOW READING
operation, where-upon all the menus become password protected (see page 21).
Serial# 0012345 V 00.00.000
Please wait.........
MAIN MENU DD-MM-YY HH:MM:SS
Quick start
View/Edit Site Data
Setup Instrument
Read flow
SETUP INSTRUMENT DD-MM-YY HH:MM:SS
*Set Date & Time : dd-mm-yy hh:mm:ss
Calibrate 4-20mA
Factory settings
Change Language
Exit